Data
75,029 bulk items are available, covering 70,621 participants.
Defined-instances run from 2 to 3, labelled using Instancing
2.
The values held within this field are coded links to files held within a bulk repository, not the actual data itself.
2 Instances
Instance 2 : Imaging visit (2014+)
70,335 participants, 70,335 items
Data not meaningful for statistical summary
Instance 3 : First repeat imaging visit (2019+)
4,694 participants, 4,694 items
Data not meaningful for statistical summary
Notes
The DICOM images were acquired using the following parameters:
Resolution: 2.4x2.4x2.4 mm
Field-of-view: 88x88x64 matrix
Duration: 4 minutes (332 timepoints)
TR: 0.735 s
TE: 39ms
GE-EPI with x8 multislice acceleration, no iPAT, flip angle 52 degrees, fat saturation
As implemented in the CMRR multiband acquisition, a separate "single-band reference scan" is also acquired. This has the same geometry (including EPI distortion) as the timeseries data, but has higher between-tissue contrast to noise, and is used as the reference scan in head motion correction and alignment to other modalities.
2 Related Data-Fields
Field ID | Description | Relationship |
20225 | Functional brain images - resting - ... | Field 20225 is the resting counterpart of task-based Current Field |
20249 | Functional brain images - task - NIFTI | Field 20249 was derived from Current Field |